Home Electronics Heating Controllers Sample Heating Power Supply HEAT1-PS

Sample Heating Power Supply HEAT1-PS PDF Print Quotation Inquiry
Sample Heating Power Supply HEAT1-PS

Description

The HEAT1-PS is the power supply designed for heating sample on sample holder under clean UHV-conditions using resistance heating method (heater is located close to the sample which becomes hot).

This method is used for temperature up to 1500 °C. Temperature depends on max power available in the version of HEAT1-PS.


Application

The HEAT1-PS is used for resistance heating method only. The Power Supply can also be used for effusion cell evaporators. The unit is equipped with PID temperature controller Eurotherm 2408 which stabilizes temperature at desired level. Ramp power heating protects the heater from overheating the sample.

The user can work in AUTO mode (with temperature control) or MANUAL mode (without temperature control).


Additional information

For fast cooling of the sample HEAT1-PS can control the cooling valve (optional). The heating process can be controlled by the computer through the RS-232 or RS-485 communication interface and dedicated application. The HEAT1-PS is mounted in the 19" rack (full-width, 3 units height) or free standing.

 

Specification
Max Power Heating Standard: 300W – r (max 600W for order)
Heater current Standard: 0 ÷ 10A at max 30V
Maximum: 0 ÷ 20A at max 30V
Temperature Range 20°C ÷ 1500°C for thermocouple type C
(dependent on sample holder type or evaporator)
– 200°C ÷ 1250°C for thermocouple K
(dependent on sample holder type and the conditions in the chamber)
Thermoelement type Standard thermocouple type K and C (other thermocouple for order)
Temperature control Process temperature control with PID Controller Eurotherm 2408
Process parameters Eurotherm 2408 finds optimized process PID parameters with autotuning
function.
Ramp heating Ramp function of Eurotherm 2408 protects the sample against overheating
Cooling down control Controls the cooling valve. Fast cooling down the sample (option).
Communication Interfaces EIA-232, 2 wire EIA-485.
Dimensions (W x H x D) 483 mm x 133 x 350 mm


 

 
EnglishPolski
This website is not an offer within the meaning of the Civil Code, and has information character only. Contact with our specialist before you order, please.