PCU14 PDF Print Quotation Inquiry
PCU14

Description

The Power Control Unit PCU14 it is the compact, stand alone microprocessor device to control one complete UHV-pumping system. Its including ultrahigh vacuum pump (e. g. turbo molecular pump, ion pump, cryo pump, etc.), electro-magnetic vent valve, electro-magnetic safety valve, fore vacuum pump and vacuum gauges based on 0..10V output standard. As standard all of Oerlikon-Leybold gauges are implemented: ITR100, ITR90, PTR237/9, TTR90 (any other can be implemented as option).
State of all peripheral devices connected to the PCU14 is currently displayed on big LCD graphics display, giving clearly readouts of basic parameters to the user.
Unit comprises several interlocks to protect the vacuum system during power failure or in the case of breakdown of its components.

 

Application

  • Any high and ultra high vacuum system
  • Measurement of ultimate pressure
  • Bakable and non bakable systems
  • Pressure measurements and control from atmosphere to UHV range

Implemented PCU14 software fully controls shutdown and venting process of theUHV-pumping system. Attached external protective battery (UPS) protect controlled system against unexpected power loses.

 

PCU14 may be equipped with RS485 or 232 / USB serial interface to enable remote control and data collecting eg. vacuum etc.

 

Additional information

PCU14 control panel is divided into two parts:
  • Center part graphical presentation of vacuum system with graphics LCD
  • Bottom part key buttons to operate PCU14

 

Mounting

PCU14 is mounted in the 19” rack (1/2 width, 3 units height). Two such units can be placed next to each other.

There are one types of PCUs but several possibilities controlled system like :
  • When the safety valve is
  • Located between fore vacuum and turbomolecular pumps
  • When the safety valve valve is between turbomolecular pump and the UHV chamber
  • Other design of vacuum system
  • Current schematic of controlled vacuum system is displayed on graphics LCD.


Specification
Measurement chanel 2
Display range 1x10-10 ÷ 1000 mbar
Display LCD 240 x 128 dots
Suitable gauges in standard* TTR211S, TTR90, PTR237, ITR90, ITR100
Dimensions [mm] ½ of 3HE 19" 212 x 128 x 235
Weight, approx 1.5 kg
Supply 90 to 240VAC
Operating temperature 5 ÷ 50°C
Interface RS232/485, USB
Supported valves any 24V 2A max
Supported fore pump any 1 phase 230VAC 5A max
or 1 - 3 phase with external interface
Supported TMP RS234/485 connected via TURBO.DRIVE S
any with simple contacts control ON/OFF
Computer interface RS232, RS485, USB

 
EnglishPolski
This website is not an offer within the meaning of the Civil Code, and has information character only. Contact with our specialist before you order, please.