MG13 PDF Print Quotation Inquiry
MG13

Description

The Multi Gauge MG13 controller is a control unit which is able to supports the operation of 2 vacuum gauges Bayard-Alpert Gauge and any others existing gauges on the market based on 0÷10V output standard. As a standard all of rlikon-Leybold gauges are implemented: ITR 100, ITR 90, DI 200, DI 2000, TTR 211 S, TTR 91, PTR 225. Any others can be implemented as an option. Inspite of 1HU unit only, controller is able to display both of channels at once on the big alphanumeric LCD display. Nice blue alphanumeric display allows to clearly readouts.

As a big advantage of this instrument it can be possibleities to use up to four triggers, which can programmable for each channel separately, with own programmable hysteretic also as 'below' or 'above' options. The measurement channels are active immediately after applying the power, and ready to emision. Channel one is dedicated to control B-Agauge only. Really flexible settings of B-Agauge parameters allows to connect almost all of B-A gauges available on the market. Such a 2 channel controller combination is perfectly suitable in 2 stages pumps group units, where two ranges (low and high vacuum) has to be controlled. More over the automatic emission control facility can be used to automatic EMISSION of B-A gauge control related to the low vacuum measurement. All operating modes of the triggers and the transmitters are displayed and their status is indicated via the outputs.


Adventages

  • 1 HU/42HP unit only in 19” Standard
  • Measurement range from atmosphere to 10-12 mbar
  • Almost all of Bayard-Alpert gauges are suitable
  • Degassing of Bayard-Alpert with power and time limit filaments supports Ir/W/Thoria
  • RS 232 or RS 485 interface in standard
  • Bayard-Alpert overpressure protection
  • Selectable measurement units: mbar, Torr, Pascal
  • Gas correction of ionization gauges possibilities, with one customized setting
  • Universal power supply: 90÷240VAC, 50÷60Hz
  • Unit overtemperature protection
  • Sound alarm
  • Four setpoints adjustable, with hysterisis with 'above' and 'below' option (individual programmed)
  • User-definied channel naming


Applications

  • Any high and ultra high vacuum systems
  • Measurement of ultimate pressure (from atmosphere to UHV range)
  • Bakeable and non bakeable systems

 

Specification
Measurement chanels 2
Mesurement range:
Chanel 1: B-A gauge
Chanel 2: 0 ÷ 10V DC input
1x10-11 mbar
5x10-10 mbar*
Display LCD 2x20 characters
9.96 mm high each line
Setpoints ( individual programming
for each channel possible )
2 (4 options)
Suitable Gauges in standard** DI 200, DI 2000,TTR 211 S,
TTR91, PTR 225,
ITR 90, ITR 100
Hysteresisadjustable
Dimentions [mm]: (˝ of 1HE) 240x260x44
Weight 1 kg
Supply 90÷240 VAC
Operating temperature 5÷50 C
Interface

RS 232 or 485 Profibus on
request

* related to the connected gauge
** any other can be implemented on request

 
EnglishPolski
This website is not an offer within the meaning of the Civil Code, and has information character only. Contact with our specialist before you order, please.
Copyright © 2009 . All Rights Reserved. PREVAC sp. z o.o.