Hemispherical Energy Analyser EA15-HP50

Technical data
Mounting flange | DN 200 CF |
Bakeout temperature | up to 120 °C |
Working distance | 0.3-2 mm (dependent on cone opening) |
Analyser mean radius | 150 mm |
Pass energies | 5, 10, 20, 50, 100, 200, 250, 300, 500 eV |
Energy resolution XPS |
< 20 meV FWHM |
Kinetic energy range | 0.5 - 1500 eV |
Acquisition modes | fixed, scan |
Transmission and angular lens mode: ▫ lens acceptance angle (transmission mode) |
+/- 9° |
Maximum energy window in fixed mode | 12.5 % of pass energy (for MCP-CCD) |
Completely designed of non-magnetic materials |
The PREVAC EA15-HP50 hemispherical energy analyser allows due to special designed extraction cones and automatic vacuum control, high-resolution PES measurements in static and dynamic environments up to 50 mbar.
The analyser is equipped with a total number of 11 slits and offers the possibility to choose between best energy resolution and best intensity. According to given photoelectron energy the analyser is set up with up to 9 predefined PE to satisfy customer's requirements.
Detectors
• MCP-CCD detector
- 40 mm diameter dual MCP detector
- 656 energy channels available simultanously
- 494 angular channels available simultanously
- 90 fps
• 7-MCD detector
The PREVAC analyser package includes:
▪ EA15-HP50 150 mm hemispherical energy analyser
▪ RUDI-EA2 high stable and low noise electronics
▪ SPECTRIUM acquisition and analyser control software for Windows OS computer system.
Additional information
EA15 hemispherical energy analyser can be controlled directly by LabVIEW programming environment.
Note:
The use of a mu-metal analysis vacuum chamber or an inner mu-metal shielded chamber in combination with PREVAC EA15-HP50 hemispherical energy analyser is recommended in order to minimalize influence of external magnetic fields and to maximize performance of the analyser.