Technical data

XREC20-I

Supply voltage 110 or 230 VAC, 50/60 Hz,
(power consumption max 1800 W)
Emission current range (Ie) 0 - 100 mA, accuracy 0.1 mA
Anode high voltage range (UHV) 0.5 kV - 15 kV, accuracy 0.1 kV
Cathode current (Icath) 3A in Standby mode
up to 20A in Operate mode
Anode power limit 650 W
High voltage ramp 1 - 1000 V/sec
Emission curent ramp 0.1 - 50 mA/sec
Interlocks High vacuum
Cooling water (master)
X-ray cover
Communication Interface RS232/485, Ethernet
Communication protocol MODBUS-TCP
XREC20-I user interface 7" TFT display with touchscreen, digital encoder
Interface languages English, German, Polish
Dimensions XREC20-I 483 x 133 x 380 mm (W x H x D),
19" rack mountable
Dimensions XRHV02-PS [mm] 483 x 88 x 420 (W x H x D), 19" rack mountable
Weight (approx.) 10 kg

XRHV02-PS

Supply voltage 85 - 260 VAC, 50-60 Hz,
Output voltage range Vo 0 - 15 kV
Output current range Io 0 - 100 mA
Max output power 1.5 kW
Polarity negative
Remote control via indirect coupled analogue I/O (male D-SUB 9)
set and monitor 0 - 5 V
Protection over load and short circuit, voltage supply and temperature
Dimensions 483 × 88 × 420 mm (W×H×D), 19″ rack mountable
Weight (approx.) 6.5 kg

XREC20 electronic set dedicated for full control of the SAX X-ray sources.

The XREC20 set comprises 2 units for full control of the SAX X-ray Sources: XREC20-I emission controller and XRHV02-PS high voltage power supply.

The XREC20-I unit controls the emission current of the X-ray source and also the XRHV02-PS high voltage power supply, displaying both the high voltage and emission current on a single display. Unit allows measurement of working time and anode power. The X-ray source electronics set has both operational and standby modes, guaranteeing the most accurate and stable emission at the start of an experiment. Easy firmware update via USB port (XREC20-I). Unit can be remotely controlled via RS232/485 or Ethernet interfaces. Autosave function (XREC20-I unit) for saving your parameters/unit preset and applying them automatically after restart.