Power supplies & heating controllers

Sample Heating Power Supply HEAT3-PS

Sample Heating Power Supply HEAT3-PS
Quotation inquiry catalog.whishlist.tooltip  

Technical data

Supply voltage 100-130 VAC/200-260 VAC, 50-60 Hz (power consumption max 1600 W)
Resistance heating mode parameters 45 V, 17 A - standard; other versions on request
Electron bombardment mode parameters 1000 V, 300 mA - standard, other versions on request
Temperature range

273.15 K-2473.15 K for type C thermocouple
(dependent on sample holder type or evaporator)


73.15 K-1645.15 K for type K thermocouple
1.4-500 K for DT670/DT470 silicon diodes
(dependent on sample holder type and conditions in chamber)

Temperature independent input 2 - for thermocouples K/C/E/N
2 - for silicon diodes DT670/DT470
Temperature setpoint ramp rate adjustable from 0.1° to 1000°/ s/min/h
ΔT setpoint 0.1 to 5.0°/s
Vacuum measurement (optional) CTR90, TTR91, TTR211, PTR225, PTR90, ITR90, ITR100, Baratron, ANALOG IN, MKS937A, PG105, MG13/14, PKR251/360/361, PCR280, ATMION
Communication interface RS232/485, Ethernet
Communication protocol MODBUS-TCP
User interface 7″ TFT display with touchscreen, digital encoder
Interface languages English, German, Polish
Dimensions [mm] 448.8 x 132.5 x 375 mm (W x H x D), 19" rack mountable
Weight (approx.) 8.8 kg (for standard version)

 

The HEAT3-PS is used for resistive heating or electron bombardment heating.

The power supply can also be used for effusion cell evaporators. The unit is equipped with a PID temperature controller.
Ramp heating function control sample temperature to protect sample from damage. Sample overheating can also be protected by setting the voltage and current limits. The unit can be operated in auto mode (with temperature control) or manual mode (without temperature control).

Features

• dual heating mode: resistive and electron bombardment
wide range temperature measurement (1.4-2473.15K)
2D real time chart module
• Shutter Control - up to 2 shutters for double DC module version, 1 shutter per channel (eg. shutters of sources or manipulators)
high efficiency
setpoint based over-voltage and over-current output protection
fully manual or PID temperature controlled (by setpoint and ramp)
process temperature control with built-in PID controller (with autotuning function for optimized process PID parameters)
various kind of temperature sensors: thermocouples K/C/E/N (other on request), Pt, diode
multiple I/O (10 digital/4 analog) - individual reprogrammable
high resolution (16-bit analog I/O, 0.1 K temp.)
one vacuum channel for active gauges
mobile solutions for remote access and control
customised menu options (for robust and effectivity)
support (easy firmware update via USB)
autosave function

Applications

any thermal process according to the specifications
effusion cells supply
sample holders heatings
thermal monitoring

Options

second DC module available:
 - two resisting heating zones with independent control, or
 - one resistive heating zone with higher output power (90V, 17A or 45V, 30A)
software control (Version Pro, Extended or Library Module)
analog I/O card for vacuum measurement (1 gauge)

Download brochure

Related products

» Components

Effusion Cell EF 40C1

» Effusion Cell EF 40C1

EF 40C1 Effusion Cell (Knudsen Cell) is a source designed for indirect resistive heating and delivering highly constant evaporation rates.

read more
Thermal Desorption Spectrometer TDS 40A1

» Thermal Desorption Spectrometer TDS 40A1

The TDS 40A1 is designed for Temperature Programmed Desorption (TPD) applications.